Drilling for implant site preparation generates heat, which can cause bone necrosis if temperatures exceed 47 °C for over a minute. Factors influencing heat include drill size, speed, pressure, irrigation, and tool wear. Frequent drill replacement is essential, as wear from repeated use and sterilization affects performance. This study compared three pilot drills with similar designs from different manufacturers, testing each on pig ribs for 15 perforations after 15 sterilization cycles. Researchers measured temperature increase, drilling time, and surface wear. Results showed that drill no. 1 generated more heat than drills no. 2 and no. 3, though none reached critical temperatures. Drill no. 2 took the longest to reach the desired depth and displayed the most deformation. Findings highlight the importance of adhering to the recommended operational limits, suggesting that drills should be replaced after 15 cycles to ensure efficacy and patient safety.

Microscopic Analysis and Evaluation of Thermal Elevation and Wear of Drills for Implant Site Preparation: An In Vitro Study

Mummolo, Stefano
2024-01-01

Abstract

Drilling for implant site preparation generates heat, which can cause bone necrosis if temperatures exceed 47 °C for over a minute. Factors influencing heat include drill size, speed, pressure, irrigation, and tool wear. Frequent drill replacement is essential, as wear from repeated use and sterilization affects performance. This study compared three pilot drills with similar designs from different manufacturers, testing each on pig ribs for 15 perforations after 15 sterilization cycles. Researchers measured temperature increase, drilling time, and surface wear. Results showed that drill no. 1 generated more heat than drills no. 2 and no. 3, though none reached critical temperatures. Drill no. 2 took the longest to reach the desired depth and displayed the most deformation. Findings highlight the importance of adhering to the recommended operational limits, suggesting that drills should be replaced after 15 cycles to ensure efficacy and patient safety.
File in questo prodotto:
File Dimensione Formato  
materials-17-05524_compressed.pdf

accesso aperto

Licenza: Creative commons
Dimensione 953.55 kB
Formato Adobe PDF
953.55 kB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11697/250039
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 0
social impact