The recoil proton energy response has been measured by exposing cylindrical cells, filled with BC501A BICRON liquid scintillator, to mono-energetic neutron reference fields. We determine the required calibration parameters and report the detailed procedures for the experimental data handling. A dedicated Monte Carlo simulation of the detector response and efficiency has been performed. It showed good agreement with the measured quantities. The results from this calibration are necessary for a detailed study of the neutron spectrum at the underground Gran Sasso Laboratory, with a neutron detector made of 32 liquid scintillator cells, like those used during the calibration. (C) 1998 Elsevier Science B.V. All rights reserved.
Titolo: | Calibration of BC501A liquid scintillator cells with monochromatic neutron beams | |
Autori: | ||
Data di pubblicazione: | 1998 | |
Rivista: | ||
Abstract: | The recoil proton energy response has been measured by exposing cylindrical cells, filled with BC501A BICRON liquid scintillator, to mono-energetic neutron reference fields. We determine the required calibration parameters and report the detailed procedures for the experimental data handling. A dedicated Monte Carlo simulation of the detector response and efficiency has been performed. It showed good agreement with the measured quantities. The results from this calibration are necessary for a detailed study of the neutron spectrum at the underground Gran Sasso Laboratory, with a neutron detector made of 32 liquid scintillator cells, like those used during the calibration. (C) 1998 Elsevier Science B.V. All rights reserved. | |
Handle: | http://hdl.handle.net/11697/6210 | |
Appare nelle tipologie: | 1.1 Articolo in rivista |